Configuration Menu (Overview)
59
EN
Configuration menu II
Menu level
1. Submenu level
2. Submenu level
3. Submenu
level
Gas flow rate*
High Flow:
Rate 1=
3 l/min
(range 1-20)
Rate 2=
20 l/min
(range rate 1-40)
Rate 3=
40 l/min
(range rate 2-40)
Venting con-
trols
Bariatric:
Rate 1=
5 l/min
(range 1-20)
Rate 2=
25 l/min
(range rate 1-45)
Rate 3=
45 l/min
(range rate 2-45)
Vessel Harvest:
Rate 1=
1 l/min
(range 1-10)
Rate 2=
4 l/min
(range rate 1-10)
Rate 3=
10 l/min
(range rate 2-10)
First nominal
gas flow
Pediatric:
0.1 l/min
(setting range 0.1-1 l/min)
Max. nominal
pressure
High Flow: 30 mm Hg
(setting range 5-30 mm Hg)
Bariatric:
30 mm Hg
(setting range 5-30 mm Hg)
Vessel Harvest:
20 mm Hg
(setting range 5-
20 mm Hg)
Pediatric:
20 mm Hg
(setting range 5-20 mm Hg)
Flow safety
limit
Limit ON
Limit OFF
Warning signal
Occlusion
Vessel Harvest: Signal ON, Signal OFF
Bariatric: Signal ON, Signal OFF
Pediatric: Signal ON, Signal OFF
High Flow: Signal ON, Signal OFF
*Inactive menu items are displayed in gray.
Menu item Gas Flow Rates is inactive in Pediatric mode.
Menu item First Nominal Gas Flow is active exclusively in Pediatric mode.
Menu item Flow Safety Limit is active only in Pediatric operating mode and Vessel Harvest operating mode.
Summary of Contents for PneumoSure
Page 6: ......
Page 101: ...Appendix 95 EN 19 Appendix 19 1 Test Log Date Results Comment Signature ...
Page 104: ......
Page 107: ...3 ES 19 Apéndice 100 19 1 Registro de test 100 Índice alfabético 101 ...
Page 108: ......
Page 204: ...100 Apéndice ES 19 Apéndice 19 1 Registro de test Fecha Resultado Anotaciones Firma ...
Page 209: ...3 FR 19 Annexe 99 19 1 Procès verbal de tests 99 Index alphabétique 100 ...
Page 210: ......
Page 305: ...Annexe 99 FR 19 Annexe 19 1 Procès verbal de tests Date Résultat Remarque Signature ...
Page 308: ......
Page 405: ...Anexo 97 PT 19 Anexo 19 1 Protocolo de teste Data Resultado Anotações Assinatura ...
Page 408: ......
Page 409: ......