Using and Controlling the PNEUMO SURE High Flow Insufflator in High Flow Mode
33
EN
6.6.4
Stop the Insufflation
1. Press the
STOP
function field. The following values are displayed:
• Gas consumption display: last measured value
• Actual pressure: current measured value
• Actual gas flow: 0 l/min
• Nominal pressure: last set value In case of exceeding the safety limit, the
nominal pressure value will be reset to the safety limit value.
• Nominal gas flow: last set value
The status field of the display depicts
Insufflation stopped
followed by
Mode:
High Flow
alternating with
Push
to release
.
2. Remove the tube set from the device. Observe applicable hygiene regulations
when disposing of the tube set.
CAUTION!
If the tube set remains connected to the device, there is the danger that leftover
fluid in the tube or the instruments will be sucked into the device.
3. Close the gas supply.
4. Use the ON/OFF switch to turn the device off.
NOTE!
Observe applicable hygiene regulations when disposing of the tube set.
Summary of Contents for PneumoSure
Page 6: ......
Page 101: ...Appendix 95 EN 19 Appendix 19 1 Test Log Date Results Comment Signature ...
Page 104: ......
Page 107: ...3 ES 19 Apéndice 100 19 1 Registro de test 100 Índice alfabético 101 ...
Page 108: ......
Page 204: ...100 Apéndice ES 19 Apéndice 19 1 Registro de test Fecha Resultado Anotaciones Firma ...
Page 209: ...3 FR 19 Annexe 99 19 1 Procès verbal de tests 99 Index alphabétique 100 ...
Page 210: ......
Page 305: ...Annexe 99 FR 19 Annexe 19 1 Procès verbal de tests Date Résultat Remarque Signature ...
Page 308: ......
Page 405: ...Anexo 97 PT 19 Anexo 19 1 Protocolo de teste Data Resultado Anotações Assinatura ...
Page 408: ......
Page 409: ......