62
Configuration Menu (Overview)
EN
For example, press the
Venting system OFF
function field.
You now have the opportunity to
1. save the selected value by pressing the
(SAVE)
function field. The display de-
picts
(SAVED)
for 2 seconds. After saving you return automatically to the pre-
vious level,
2. press the
function field to return to the previous menu level without sav-
ing,
3. or press
(EXIT)
to return to the work screen without saving.
Setting the venting pressure limit of the
venting valve
Factory setting = 3 mm Hg
(for all operating modes)
For example, press the
Venting pressure limit
function field.
The display switches to the
Pressure limit
selection menu.
You now have the opportunity to
1. save the selected value by pressing the
(SAVE)
function field. The display de-
picts
(SAVED)
for 2 seconds. After saving you return automatically to the pre-
vious level,
2. press the
function field to return to the previous menu level without sav-
ing,
3. or press
(EXIT)
to return to the work screen without saving.
Setting the venting system response time
Factory setting = 3 s
(for all operating modes)
For example, press the
Venting response time
function field.
The display switches to the
Venting response time
selection menu.
For example, press the
4 s
function field.
You now have the opportunity to
1. save the selected value by pressing the
(SAVE)
function field. The display de-
picts
(SAVED)
for 2 seconds. After saving you return automatically to the pre-
vious level,
2. press the
function field to return to the previous menu level without sav-
ing,
3. or press
(EXIT)
to return to the work screen without saving.
10.1.3 Setting the Gas Supply Type
Use this menu to select the type of connected gas supply.
Select
House Gas
if you are working with a house gas supply (use the correspond-
ing house gas device adapter available as an additional accessory). The gas supply
(1)
(3)
(2)
(1)
(3)
(2)
(1)
(3)
(2)
Summary of Contents for PneumoSure
Page 6: ......
Page 101: ...Appendix 95 EN 19 Appendix 19 1 Test Log Date Results Comment Signature ...
Page 104: ......
Page 107: ...3 ES 19 Apéndice 100 19 1 Registro de test 100 Índice alfabético 101 ...
Page 108: ......
Page 204: ...100 Apéndice ES 19 Apéndice 19 1 Registro de test Fecha Resultado Anotaciones Firma ...
Page 209: ...3 FR 19 Annexe 99 19 1 Procès verbal de tests 99 Index alphabétique 100 ...
Page 210: ......
Page 305: ...Annexe 99 FR 19 Annexe 19 1 Procès verbal de tests Date Résultat Remarque Signature ...
Page 308: ......
Page 405: ...Anexo 97 PT 19 Anexo 19 1 Protocolo de teste Data Resultado Anotações Assinatura ...
Page 408: ......
Page 409: ......