Error and Warning Messages
EN
Gas temperature >42°C/ Disconnect
luer lock / Cool down tube
The device terminates gas heating
and insufflation.
The temperature of the gas
exceeds 42° C.
• Unplug the insufflation tube at the trocar.
• Press the
START/STOP
function field. The device
insufflates without heating the gas.
• Let hot gas escape until the tube is only warm to
the touch.
• Continue the operation without gas heating or use
a different gas heating tube.
Turn device off and on again. This reactivates the
gas heating.
• Check gas heating after surgery using a different
tube.
If this error message is displayed again, have the
device checked by a qualified service technician.
It is possible to continue using the device without gas
heating.
Error message/Call service
The device does not work prop-
erly and activates the internal
safety system.
• Switch the device off and back on after approx. 3
seconds have expired using the ON/OFF key.
• Should the error message appear again, the device
is defective. Make sure the device can no longer be
used until a qualified service technician conducts
the appropriate tests and repairs.
Device temperature error/Turn
device off
The insufflation is stopped or cannot
be started.
The device temperature is
above 70 °C.
• Prepare for changing the gas bottle.
The device temperature is
below 10° C.
• Turn the device off for about 10 minutes using the
ON/OFF key. The surrounding room temperature
must be above 10° C.
Venting valve defective
Venting system malfunctioning. • Use the ON/OFF key to turn the device off and turn
device back on after approx. 3 seconds. If this error
message is displayed again after the device check
has concluded, have the device checked by an
authorized service technician. The device can be
still operated but without venting system. The error
message is repeated with each Start/Stop.
RTP defective
Device defective
• It is possible to continue with the surgery but
do
not connect the measuring line of the RTP system
.
The message is repeated with each Start/Stop.
After surgery is finished, call a service technician to
repair the device.
RTP deactivated
Device check was unsuccessful
• Surgery can be continued but without the continu-
ous pressure measurement. Please make sure that
neither of the two tubes is connected to a trocar or
closed off during a device check. Restart device if
needed.
Tube set not connected
The tube set has not been con-
nected or has not been inserted
correctly but the START function
field has been pressed.
• Please insert a tube set or check the tube set con-
nection.
Check/change tube set
Leak in tube connection
• Please insert a new tube set or check the tube set
connection.
Error and Warning Messages
Cause
Troubleshooting
Summary of Contents for PneumoSure
Page 6: ......
Page 101: ...Appendix 95 EN 19 Appendix 19 1 Test Log Date Results Comment Signature ...
Page 104: ......
Page 107: ...3 ES 19 Apéndice 100 19 1 Registro de test 100 Índice alfabético 101 ...
Page 108: ......
Page 204: ...100 Apéndice ES 19 Apéndice 19 1 Registro de test Fecha Resultado Anotaciones Firma ...
Page 209: ...3 FR 19 Annexe 99 19 1 Procès verbal de tests 99 Index alphabétique 100 ...
Page 210: ......
Page 305: ...Annexe 99 FR 19 Annexe 19 1 Procès verbal de tests Date Résultat Remarque Signature ...
Page 308: ......
Page 405: ...Anexo 97 PT 19 Anexo 19 1 Protocolo de teste Data Resultado Anotações Assinatura ...
Page 408: ......
Page 409: ......