60
Configuration Menu (Overview)
EN
10.1
Configuration menu I
Functions of the configuration menu
While insufflation is stopped, press the
Actual
gas flow function field for 2 sec-
onds to access the Configuration menu I
10.1.1 Setting First Nominal Pressure
Utility menu
Menu level
1. Submenu level
2. Submenu level
3. Submenu level
Display setting
Dimmer
Dimmer sensor
-75%
, -50%, -25%
Activated
OFF
Language
english
français
deutsch
español
português
italiano
nederlands
norsk
suomi
greek
svenska
dansk
polski
românã
simplified chi-
nese
korean
japanese
Program version
Upgrade XL
Upgrade from version Pneumo Sure to Pneumo Sure XL
Service
Access only for service technicians
*Inactive menu items are displayed in gray.
Depiction/functions in configuration menu
Tap the "arrow forward" function field to access the next
menu on the same level.
Tap the "arrow backward" function field to access the
previous menu on the same level
without saving set-
tings
or to access the previous menu of the next higher
menu level.
Tap the
(SAVE)
function field to save settings. The display
depicts
(SAVED)
for 2 seconds. After saving you return
automatically to the previous level.
Press the
(EXIT)
function field to exit the menu or return
to the work screen.
Operating mode
Factory setting
Range
Bariatric
15 mm Hg
1-15 mm Hg
High Flow
15 mm Hg
1-15 mm Hg
Vessel Harvest
10 mm Hg
1-12 mm Hg
Pediatric
8 mm Hg
1-12 mm Hg
START
Set Pressure
Actual Pressure
mm Hg
mm Hg
Set Flow
l/min
Mode: High Flow
3
15
Change
Mode
Actual
l/min
0
Liter
0.0
0
GAS
(1)
Summary of Contents for PneumoSure
Page 6: ......
Page 101: ...Appendix 95 EN 19 Appendix 19 1 Test Log Date Results Comment Signature ...
Page 104: ......
Page 107: ...3 ES 19 Apéndice 100 19 1 Registro de test 100 Índice alfabético 101 ...
Page 108: ......
Page 204: ...100 Apéndice ES 19 Apéndice 19 1 Registro de test Fecha Resultado Anotaciones Firma ...
Page 209: ...3 FR 19 Annexe 99 19 1 Procès verbal de tests 99 Index alphabétique 100 ...
Page 210: ......
Page 305: ...Annexe 99 FR 19 Annexe 19 1 Procès verbal de tests Date Résultat Remarque Signature ...
Page 308: ......
Page 405: ...Anexo 97 PT 19 Anexo 19 1 Protocolo de teste Data Resultado Anotações Assinatura ...
Page 408: ......
Page 409: ......