10 Alarms and troubleshooting
ASTOFLO PLUS ECO Instructions for Use
41
10.5 Connection alarm
Control panel
Device response
The display
indicates
C0
.
The “Start” LED
flashes.
The “Alarm” LED
turns on.
There is a brief acoustic alarm signal every 16
seconds.
Alarm condition
The heating profile is disconnected from the control unit
in the
Heating Mode
or no heating profile is connected
and the “Start” button
is pressed in
On Mode
.
Possible reason(s)
►
Required action(s)
No heating profile is connected or heating profile
became disconnected.
►
1. Connect the heating profile to the control unit.
2.
Press the “Start” button
to switch the device to
Heating Mode
.
Possible actions to
clear
-