12 Maintenance
ASTOFLO PLUS ECO Instructions for Use
45
12 Maintenance
ASTOFLO PLUS ECO does not require preventive maintenance (e.g. replace-
ment of liquids or components). Recurrent tests shall be performed according to
chapter 12.1.
No service or maintenance shall be carried out while in use with a
patient.
WARNING
Risk of injury!
The service personnel must be appropriately trained and qualified.
The ASTOFLO PLUS ECO does not contain any parts the user can repair.
Therefore, do not attempt to repair the ASTOFLO PLUS ECO yourself.
Contact your local sales point.
Any repairs (such as, but not limited to, changing the power supply cord) to
the equipment may only be carried out by persons authorised and qualified
by the manufacturer.
Modifications to the device are not permitted.
The accessories specified in chapter 15 may be replaced without restriction by
the operating or maintenance personnel.
On request, STIHLER ELECTRONIC GmbH will provide service instructions that
will allow properly trained and qualified persons to repair the parts of the
equipment that the manufacturer has designated as repairable.
Provision of technical documents and/or spare parts is not an authorisation from
the manufacturer to open or repair the equipment.
12.1 Recurrent tests
12.1.1 Control unit
(Heating profile see 12.1.2)
A recurrent test must be carried out on the ASTOFLO PLUS ECO control unit at
least every 12 months to ensure the safe operation of the warmer.
Please ensure that all the applicable national directives (e.g. IEC/EN 62353) for
checking the safety of medical equipment are observed additionally and that the
test equipment is calibrated.
Necessary test equipment:
Standard medical electrical safety tester
Room thermometer
Stopwatch
The following sections describe how the test is to be performed, and the modes of
operation for the recurrent test are defined. The attached test protocol
(chapter
can be used.