ASTOFLO PLUS ECO Instructions for Use
12 Maintenance
54
Test 4.2: Applied part leakage current (alternative method)
Optional to test 4.1
Min [mA]
Value [mA] Max [mA] P/F
Applied part leakage current
0.005
0.05
Manual tests
P/F
Test 5:
Manual excessive temperature cut off Test (E11, E12)
Test 6:
Manual cable break cut off test
Heating profile
P/F
Test 7:
Visual inspection (damages, marking)
Test 8:
Temperature sensors heating profile
Value [°C]
Max [°C] P/F
Temperature sensor 1 (T1)
Temperature sensor 2 (T2)
Temperature thermometer (TT)
Difference TT to T1
1.2
Difference TT to T2
1.2
Difference T1 to T2
1.2
Test 9:
Heating-up test
Min
Value
Max
P/F
Heating up time
min.
150 s
Fluctuation of temperature display
42.8°C
-
°C
43.2°C
Overall assessment
Please mark where applicable
No safety or functional deficiencies were detected
No direct risk, deficiencies detected may be corrected on short term
Equipment must be taken out of operation until deficiencies are corrected
Equipment does not comply
– Modification / Exchange of components / Taking out
of service - is recommended
Comments
Date
Signature