6 Product description
20
Fig. 5 ASTOFLO PLUS ECO back
#
Part
Description
7
Universal Attachment
Device
For secure attachment of the control unit.
8
Clip for Heating Profile
Fixes the heating profile/infusion line.
9
Knurled screw
Protects the device against accidental detaching from
the medical rail.
10
Connector for Potential
Equalization
The purpose of additional potential equalization is to
equalize potentials between different metal parts that
can be touched simultaneously, or to reduce differences
of potential which can occur during operation between
the body, the medical electrical devices and conductive
parts of other objects.
The connection is made by green-yellow insulated leads
(min. 4 mm²) to standardized plug connectors and
receptacles.
When connecting/combining medical electrical
equipment together to a medical electrical system, the
requirements of IEC/EN 60601-1 must be observed.
9
8
7