9 Getting started
ASTOFLO PLUS ECO Instructions for Use
29
9 Getting started
Getting started is grouped into 4 sections. Read through each section before
performing a procedure.
To achieve the greatest possible benefit the ASTOFLO PLUS ECO
should be set up close enough to the patient for the end of the flexible
heating profile to reach the point of injection.
The ASTOFLO PLUS ECO should be switched to
Heating Mode
before use so that the heating profile can heat up. The heating-up
process takes about 4 minutes at 20°C room temperature.
Do not position the device in a manner that is difficult to disconnect it
from mains by the mains plug.
9.1 Preparation for use
WARNING
Risk of injury!
In the event that any of the following has occurred, do not use the ASTOFLO
PLUS ECO until the appropriate corrective measures have been taken:
- Damaged or worn cables, plugs, or connecting socket.
- Damaged housing, damaged or loose control panel.
- Control unit has been exposed to mechanical impact / exposed to severe
shock or exposed to liquid.
- Alarm without knowing the cause.
- Damaged heating profile, e.g. caused by clamps, scissors or improper
handling or storage.
- Damaged or missing markings/safety signs/warnings on the control unit
and/or heating profile.
Use of the ASTOFLO PLUS ECO must be carried out under the supervision
of a physician.
The mains cable should not touch the patient and should not hinder the
treating personnel.
CAUTION
Risk of injury!
When fixing the Warmer to a mounting device, pay attention to the max. load to
avoid tilting. Using normal infusion stands ASTOFLO ECO PLUS may be
mounted at a height of up to 165 cm. If you use the robust IV pole
ASTOSTAND, the device can be mounted at a height of 2 m.