ASTOFLO PLUS ECO Instructions for Use
10 Alarms and troubleshooting
42
10.6 Standby mode - failure
Control panel
Device response
The
“Standby” LED
LED is off, and the device
cannot be switched to
On Mode
by pressing the
“Standby” button
.
Possible reason(s)
►
Required action(s)
Power supply problem or no power.
►
Check plugs and fuse, compare mains voltage with
rating plate.
Control unit not plugged in.
►
1. Plug the control unit into functioning socket.
2.
Press the “Standby” button
.
3.
Press the “Start” button
to switch the control
unit to
Heating Mode
.
Control unit is defective.
►
Return the control unit to local sales office.
10.7 On mode - failure
Control panel
Device response
The
“Standby” LED
is on, but the device cannot be
switched to
On Mode
by pressing the
“Standby” button
.
Possible reason(s)
►
Required action(s)
Control unit is defective.
►
Return the control unit to local sales office.