ASTOFLO PLUS ECO Instructions for Use
10 Alarms and troubleshooting
40
10.4 Self-test alarm
Control panel
Device response
The display
indicates
E
.
The “Alarm” LED
turns on.
There is a brief acoustic alarm signal every 16
seconds.
The control unit cannot be started.
Alarm condition
Electronics failure detected when the control unit is
switched from
Standby
to
On Mode
or from
On
to
Heating Mode
or the “Start” button
has been
pressed to soon after switching to
On Mode
.
Possible reason(s)
►
Required action(s)
“Start” button
has been pressed too soon.
►
Repeat the procedure after clearing the alarm.
Failure of the electronics of the control unit.
►
Return the Warmer to local sales office.
Possible actions to
clear
Press the
“Standby” button
to switch the device to
Standby Mode
.
7
5
4
4
2
1
3
6
5
7
4
2
8