ASTOFLO PLUS ECO Instructions for Use
9 Getting started
30
1. Attach the control unit to infusion stand or to a medical standard rail using the
attachment device according to
chapter 8.2 Installation of the warmer
2. Plug the mains plug of the control unit into a socket (
“Standby” LED turns on to
show that the control unit is in
Standby Mode
).
3. Connect the heating profile to the control unit according Fig. 7.
4. Press the
“Standby” button
to switch the warmer to
On Mode
.
WARNING
Risk of injury!
If the yellow "Alarm
” LED and the acoustic alarm are not activated
automatically when the device turns on by pressing the "Standby
” button,
remove the device from service immediately.
5. Check the acoustic and visual signals and the display:
All segments of the display and all LEDs flash and the acoustic alarm
signal beeps once to indicate that the control unit is functioning correctly.
The
“Start” LED
flashes and the temperature display shows the actual set
temperature flashing for about 3 seconds.
Then the current temperature of the heating profile is indicated.
6. Press the
“Set” button
to select another set temperature if needed.
7. Press the
“Start” button
to switch the ASTOFLO PLUS ECO to
Heating
Mode
(
“Start” LED
illuminates).
As long as the temperature of the heating profile is below 18 °C, the
display indicates low by displaying
“L”.
The temperature of the heating profile can be changed during
operation at any time (
chapter 7 Operating states - section 7.4
The displayed temperature shows the temperature of the Heating
Profile.
This is not equal to the blood
or patient’s temperature!
1.
2.
Fig. 7 Connecting heating profile