9 Getting started
ASTOFLO PLUS ECO Instructions for Use
35
Control Unit
Clean and wipe-disinfect the control unit in accordance with the procedure below:
1. Disconnect the mains plug from the socket.
2. Clean all surfaces including with a soft cloth/cotton swab and mild soap-and-
water solution.
3. Disinfect the
control unit
with either:
An approved disinfectant
An alcohol-based disinfectant with a low content of aldehydes (< 0.2%)
A mild bleach solution (max. 0.25 % hypochlorite)
Heating Profile
Clean and wipe-disinfect the heating profile in accordance with the procedure
below:
1. Clean all surfaces of the heating profile including the groove with a soft cloth/
cotton swab and a mild soap-and-water solution or with water only.
2. Disinfect the
heating profile
only with approved disinfectants or with
disinfectants based on alcohol with a low content of aldehyds (< 0.2 %).
Do not
disinfect the heating profile with disinfectants containing hypochlorite
(bleach).
In doing so follow the contact time of the disinfectant given in the specific
instructions for use of the disinfectant! After this time dry the heating profile.
3. Residuals of disinfectants cause a sticky surface, it is strongly recommended
to rinse with water after about 5 disinfections or once a week.
To facilitate insertion of the infusion line, the heating profile can be
dusted with usual in trade powder or talcum powder.
List of approved disinfectants
*
:
Meliseptol
®
Clinell Alcohol Wipes
Biguamed
®
Perfekt N
Incidin
®
Plus
Mikrozid
®
Liquid
HyPro medical 3% H
2
O
2
Bacillol
®
Plus
Aniosurf
Mikrobac
®
forte
Oxivir Tb
ClearSurf
®
Diosol 3% H
2
O
2
PURE
Clinell Universal Sanitising Wipes
Virox5 RTU
*In the United States please use only disinfectants which are registered by EPA
(U.S. Environmental Protection Agency) or cleared by FDA (U.S. Food & Drug Administration).