ASTOFLO PLUS ECO Instructions for Use
16 Guidelines and manufacturer's declaration
58
16 Guidelines and manufacturer's declaration
Guidance and manufacturer's declaration - electromagnetic emissions
The ASTOFLO PLUS ECO is intended for use in the electromagnetic environment specified below.
The customer or user of the ASTOFLO PLUS ECO should assure that it is used in such an environment.
Emission test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11/EN 55011
Group 1
The ASTOFLO PLUS ECO uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are not
likely to cause interference in nearby electronic equipment.
RF emissions
CISPR 11/EN 55011
Class A
The emissions characteristics of this equipment make it suitable for use
in industrial areas and hospitals (CISPR 11 class A). If it is used in a
residential environment (for which CISPR 11 class B is normally
required) this equipment might not offer adequate protection to radio-
frequency communication services. The user might need to take
mitigation measures, such as relocating or re-orienting the equipment.
Harmonic emissions
IEC/EN 61000-3-2
Class A
Voltage fluctuations /
flicker emissions
IEC/EN 61000-3-3
Complies
Guidance and manufacturer's declaration - electromagnetic immunity
The ASTOFLO PLUS ECO is intended for use in the electromagnetic environment specified below.
The customer or user of the ASTOFLO PLUS ECO should assure that it is used in such an environment.
Immunity test
Test level
Compliance
level
Electromagnetic environment -
guidance
Electrostatic
discharge (ESD)
IEC/EN 61000-4-2
± 8 kV contact
± 2 kV, ± 4 kV, ± 8
kV, ± 15 kV air
in compliance
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30%.
Electrical fast
transient/burst
IEC/EN 61000-4-4
± 2 kV
100 kHz repetition
frequency
in compliance
Mains power quality should be that
of a typical commercial or hospital
environment.
Surge
IEC/EN 61000-4-5
± 0.5 kV, ± 1 kV
Line-to-line
± 0.5 kV, ± 1 kV, ±
2 kV
Line-to-ground
in compliance
Mains power quality should be that
of a typical commercial or hospital
environment.
Voltage dips
IEC/EN 61000-4-11
0 % U
T
; 0.5 cycle
At 0°, 45°, 90°, 135°,
180°, 225°, 270° and
315°
0 % U
T
; 1 cycle
and
70 % U
T
; 25/30
cycles
Single phase: at 0°
in compliance
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the
ASTOFLO PLUS ECO requires
continued operation during power
mains interruptions, it is
recommended that the ASTOFLO
PLUS ECO be powered from an
uninterruptible power supply or a
battery.
Voltage interruptions
IEC/EN 61000-4-11
0 % U
T
; 250/300
cycle
in compliance
Rated power frequency
magnetic fields
IEC/EN 61000-4-8
30 A/m
50 Hz or 60 Hz
in compliance
Power frequency magnetic fields
should be at levels characteristic
of a typical commercial or hospital
environment.
NOTE UT is the AC mains voltage prior to application of the test level.