12 Maintenance
ASTOFLO PLUS ECO Instructions for Use
49
The test is not successful if any of the following conditions occurs:
The displayed temperature does not rise.
The display does not indicate
C0
.
The “Start” LED does not flash.
The “Alarm” LED does not turn on.
The acoustic alarm signal does not beep.
12.1.2 Heating profile
In order to ensure the safe operating condition, the recurrent test of the heating
profile must be carried out at least every 12 months.
Test 7
Visual inspection
Required
actions
1. Clean the heating profile with an alcohol-based agent.
2. Dust the heating profile (groove and outer side) with talcum
powder.
3. Pull the entire silicone jacket through the hand and pay
attention to:
-
unusual discolorations in the groove or on the outer
surface of the profile
-
damages, scratches, cuts or open areas in the silicone
profile
4. Check marking and the safety signs.
Result
The test is successful, when
no discolorations are visible
no damages are present
the safety signs are complete and legible