4-3
Operating Instructions
Operator Manual
150830-353
The tabletop may be articulated within the limits shown by use of the hand
control positioning buttons or the optional foot control positioning pedals, or by
the optional ORCS System. If these controls fail to function, refer to
S
ECTION
7,
T
ROUBLESHOOTING
,
to see if the problem can be quickly determined and corrected.
If problem is not readily apparent, table may continue to be operated per
procedures outlined in
S
ECTION
5, A
UXILIARY
O
VERRIDE
S
YSTEMS
. Headrest position
and kidney bridge elevation must be adjusted manually as outlined later in this
section.
NOTE: Battery-powered tables should be switched OFF after each procedure to
prevent unnecessary battery discharge. If low battery condition is indicated by
the hand control LED, refer to S
ECTION
6 for Battery Charging Procedure.
The table will continue to function normally for
at least 24 hours
after the
BATTERY DOWN LED (see S
ECTION
1, D
EFINITION
OF
S
YMBOLS
) first illuminates. If
the LED illuminates during a procedure, complete the procedure and recharge the
batteries at the end of the day. If the BATTERY DOWN LED is flashing,
immediately connect the ac power cord to the table base and plug into an
appropriate acreceptacle (see Figure 6-3).
NOTE: When quickly articulating the table through various repeated movements,
the synchronization of the tabletop sections can get out of alignment and the
hand control may stop functioning properly. To prevent this condition, the user
should perform a return-to-level function after
each
procedure. This will keep the
variance over multiple articulations to a minimum.
4.2 Tabletop
Positioning
WARNING – PINCHING
HAZARD: Pinch points are
created during extreme table-
top articulation. Carefully re-
view illustrations in Figure 2-1
before operating the table.
WARNING – PINCHING AND
TIPPING HAZARD: Patient
injury may result if the opera-
tor of this table is not com-
pletely familiar with the con-
trols for patient positioning
and table operation.
WARNING – PERSONAL
INJURY HAZARD:
•
Healthcare professionals
must ensure patients are po-
sitioned and monitored to
prevent compromising res-
piration, nerve pathways, or
circulation.
•
Unanticipated table move-
ment could cause patient in-
jury. Patient must be secured
to the table in accordance
with recommended position-
ing practices.
CAUTION – POSSIBLE
EQUIPMENT DAMAGE:
•
Hang hand control from side
rail (or end rail) of table when
not in use to avoid possible
damage to the control.
•
Route the hand control cord
(and optional HERMES
®
-
Ready
1
or ACT Enabled™ in-
terface cord and/or optional
foot control cord, if appli-
cable) clear of any pinch
points where the cord(s)
could be damaged.
4.2.1 Hand Control
Operation
NOTE: Refer to S
ECTION
7, T
ROUBLESHOOTING
, to identify problems as indicated by
red LEDs on the hand control.
The following functions must be completed before any position-
ing functions are operable:
•
Control turned ON.
•
Floor locks engaged.
1
HERMES-Ready is a registered trademark of Computer Motion.
Summary of Contents for Amsco 3085 SP
Page 1: ...MAINTENANCE MANUAL Amsco 3085 SP Surgical Table After S N B420702 025 10 07 10 P764332 897 ...
Page 3: ...OPERATOR MANUAL Amsco 3085 SP Surgical Table 08 10 11 P150830 353 ...
Page 56: ......
Page 132: ......
Page 155: ...18 764332 899 Figure 5 AC Plate Assembly 1 7 3 4 5 6 2 9 10 P17 ...
Page 161: ...24 764332 899 Figure 8 Column Assembly 1 2 3 4 5 6 7 8 9 ...
Page 176: ...40 764332 899 Figure 15 Leg Section Assemblies 1 7 2 4 3 5 6 ...
Page 204: ......
Page 205: ......
Page 206: ......
Page 207: ......
Page 208: ......
Page 209: ......
Page 210: ......
Page 211: ......
Page 212: ......
Page 213: ......
Page 214: ......
Page 215: ......
Page 216: ......
Page 217: ......
Page 218: ......
Page 219: ......
Page 220: ......
Page 221: ......
Page 222: ......
Page 223: ......
Page 224: ......
Page 225: ......
Page 226: ......
Page 227: ......
Page 228: ......