3-1
Installation Instructions
Operator Manual
150830-353
3
INSTALLATION INSTRUCTIONS
NOTE: Patient grounding post/potential equalization terminal (male connector,
DIN 42801) is provided. Mating female connector is not furnished by STERIS.
IMPORTANT:
Before connecting the table to your ac power system, check
that table internal voltage switches are set for your power system (100, 120, 220,
or 230/240).
IMPORTANT:
Battery powered tables should be completely charged prior to
initial operation. Charge batteries as indicated in
S
ECTION
6, R
OUTINE
M
AINTENANCE
,
before proceeding.
If table is to be placed in extended storage, have table prepared for storage by
a qualified service technician. Ensure batteries are disconnected and check
batteries before reconnecting. Every six months the table must be operated
through all articulations and the batteries charged.
Ensure the bellows located beneath the saddle is intact, in place, and is free of
rips, tears, and punctures.
WARNING – PERSONAL
INJURY HAZARD: If the integ-
rity of the external protective
earth conductor installation
or arrangement is in doubt,
operate the table from its in-
ternal power source.
WARNING – EXPLOSION
HAZARD: Table must not be
used in the presence of flam-
mable anesthetics.
WARNING – TRIPPING
HAZARD: Route power cord
to receptacle in a position so
it will not be tripped over by
personnel in the area.
CAUTION – POSSIBLE
EQUIPMENT DAMAGE:
•
When moving the table to
point of use, roll it carefully
at moderate speed and only
over smooth floors. Maxi-
mum floor clearance is 1/4"
(6 mm). Avoid door jambs,
elevator jambs, and obstruc-
tions greater than 1/4" (6 mm).
If necessary, lift uncrated
table over obstructions, onto
trucks, etc. Lift table evenly
and only by the table base.
DO NOT transport articles (in-
cluding accessories) on top
of the table and DO NOT use
a forklift to move the uncrated
table.
•
The table has internal
switches for setting to vari-
ous ac input voltages. Im-
proper setting of switches
may damage table electrical
system and/or cause im-
proper operation of the table.
Summary of Contents for Amsco 3085 SP
Page 1: ...MAINTENANCE MANUAL Amsco 3085 SP Surgical Table After S N B420702 025 10 07 10 P764332 897 ...
Page 3: ...OPERATOR MANUAL Amsco 3085 SP Surgical Table 08 10 11 P150830 353 ...
Page 56: ......
Page 132: ......
Page 155: ...18 764332 899 Figure 5 AC Plate Assembly 1 7 3 4 5 6 2 9 10 P17 ...
Page 161: ...24 764332 899 Figure 8 Column Assembly 1 2 3 4 5 6 7 8 9 ...
Page 176: ...40 764332 899 Figure 15 Leg Section Assemblies 1 7 2 4 3 5 6 ...
Page 204: ......
Page 205: ......
Page 206: ......
Page 207: ......
Page 208: ......
Page 209: ......
Page 210: ......
Page 211: ......
Page 212: ......
Page 213: ......
Page 214: ......
Page 215: ......
Page 216: ......
Page 217: ......
Page 218: ......
Page 219: ......
Page 220: ......
Page 221: ......
Page 222: ......
Page 223: ......
Page 224: ......
Page 225: ......
Page 226: ......
Page 227: ......
Page 228: ......