4-12
150830-353
Operator Manual
Operating Instructions
TREN
D
HEIG
HT
SIDE
TILT
BACK
LEG
Perform the following for each X-ray top section:
1. Loosen screws securing spring-loaded spacers to X-ray top section. Position
section on table.
2. Rotate spacers so spring clips are in line when viewed from beneath tabletop
(see Figure 4-13).
3. Shift X-ray top section until mounting screw shaft on one of spring-loaded
spacers is centered in hole in X-ray top section. (A 1/16" [1.6 mm] clearance
is provided between each screw shaft and its X-ray mounting hole.)
4. Hold spacer to prevent it from rotating and tighten screw.
5. Ensure remaining spring-loaded spacer is flat against tabletop and center it
in its tabletop mounting hole, then tighten screw.
NOTE: When removing an X-ray top section, grasp it at the corners where the
spring-loaded spacers are located and lift straight up. Grasping it at the
opposite corners will cause the spacers to bind. Repeat the preceding
procedure if the spacers bind when section is lifted correctly.
6. Remove replace X-ray top section several times; X-ray top should lift freely.
1. To install, place clamp (or socket) on side rail and lock in position with knob
(or handle) provided (see Figure 4-14).
2. To remove, loosen knob (or handle) and slide clamp (or socket) along side
rail until a notch is reached, then remove clamp (socket).
NOTE: Clamp (socket) may also be removed from end of side rail by raising
gravity stops.
The following special accessories have been developed specifically for use with
Typical X-Ray Top Section
Figure 4-13. X-Ray Top
Spring Clips and Standoffs
4.6.3 General Accessories
Applied to Side Rails
WARNING – PERSONAL
INJURY HAZARD:
•
When installing any table ac-
cessory, check for correct at-
tachment and tighten se-
curely (if appropriate). Do not
use worn or damaged acces-
sory. Check installation be-
fore using any accessory.
•
There is a 1,000-lb (452-kg)
patient weight limit if patient
is in normal orientation and
a 500-lb (226-kg) patient
weight limit if patient is in
reversed orientation; how-
ever, the accessory load rat-
ing may be lower. Do not
exceed the accessory load
rating if it is lower than the
table rating.
Figure 4-14. General Accessories Applied to Side Rails
Short Standoff
Spring Clip Projecting
Beyond Standoff
Short Standoff
Clamp
Socket
Summary of Contents for Amsco 3085 SP
Page 1: ...MAINTENANCE MANUAL Amsco 3085 SP Surgical Table After S N B420702 025 10 07 10 P764332 897 ...
Page 3: ...OPERATOR MANUAL Amsco 3085 SP Surgical Table 08 10 11 P150830 353 ...
Page 56: ......
Page 132: ......
Page 155: ...18 764332 899 Figure 5 AC Plate Assembly 1 7 3 4 5 6 2 9 10 P17 ...
Page 161: ...24 764332 899 Figure 8 Column Assembly 1 2 3 4 5 6 7 8 9 ...
Page 176: ...40 764332 899 Figure 15 Leg Section Assemblies 1 7 2 4 3 5 6 ...
Page 204: ......
Page 205: ......
Page 206: ......
Page 207: ......
Page 208: ......
Page 209: ......
Page 210: ......
Page 211: ......
Page 212: ......
Page 213: ......
Page 214: ......
Page 215: ......
Page 216: ......
Page 217: ......
Page 218: ......
Page 219: ......
Page 220: ......
Page 221: ......
Page 222: ......
Page 223: ......
Page 224: ......
Page 225: ......
Page 226: ......
Page 227: ......
Page 228: ......