6-4
150830-353
Operator Manual
Routine Maintenance
3. Raise table to maximum elevation to access lower surfaces.
4. Clean bellows, shrouds, and entire base surface as follows:
a. Care should be taken so bellows is not torn, punctured, or ripped, and that
seals at top and bottom are not broken.
b. Holding can 6-8" (150-200 mm) from surface, spray cleaning fluid liberally
on column skirt, cap, and shrouds.
c. Gently wipe sprayed surfaces with a clean cloth dampened with water to
remove cleaning fluid. (Dampening cloth will minimize streaking.)
d. Wipe cleaned surfaces again with a clean, damp, lint-free cloth to remove
any remaining residue.
e. Repeat
Steps b, c, and d
for base surface.
5. Turn control OFF when finished with cleaning tabletop and base.
6. Clean hand control as follows:
a. Disconnect hand control from table.
b. Holding can 6-8" (150-200 mm) from surface, spray cleaning fluid liberally
on hand control and cord.
c. Wipe sprayed surface with a clean cloth dampened with water to remove
cleaning fluid.
d. Wipe cleaned surfaces again with a clean, damp, lint-free cloth to remove
any remaining residue.
e. Reconnect hand control to table, and store by attaching to table side rail.
Perform all steps listed in
S
ECTION
6.2.1, P
OST
-U
SAGE
, cleaning procedure.
1. Perform
Steps 1 through 4
under
S
ECTION
6.2.1, P
OST
-U
SAGE
,
cleaning
procedure.
2. Articulate table through all movements and clean all additional exposed
surfaces during these articulations as follows:
a. Holding can 6-8" (150-200 mm) from surface, spray cleaning fluid liberally
on surface to be cleaned.
b. Wipe sprayed surfaces with a clean cloth dampened with water to remove
cleaning fluid. (Dampening cloth will minimize streaking.)
c. Wipe cleaned surfaces again with a clean, damp, lint-free cloth to remove
any remaining residue.
3. Turn control OFF when finished with cleaning procedures.
4. Inspect bellows for rips, tears or punctures. A damaged bellows can allow
fluids to settle into table interior.
CAUTION – POSSIBLE
EQUIPMENT DAMAGE:
•
Cleaning procedures requir-
ing articulation of the table
should be performed only by
persons familiar with table
operation.
•
Do not spray cleaning fluid
into electric receptacles and
avoid spraying directly on
override switches or into
clearance space above col-
umn. Spray or drippage may
settle onto electric circuits
inside table causing corro-
sion and loss of function.
CAUTION – POSSIBLE
EQUIPMENT DAMAGE: Hang
the hand control from side
rail (or end rail) of the table
when not in use to avoid pos-
sible damage to the control.
6.2.2 End-of-Day
6.2.3 Weekly
CAUTION – POSSIBLE
EQUIPMENT DAMAGE:
•
Cleaning procedures requir-
ing articulation of the table
should be performed only by
persons familiar with table
operation.
•
Do not spray cleaning fluid
into electric receptacles and
avoid spraying directly on
override switches or into
clearance space above col-
umn. Spray or drippage may
settle onto electric circuits
inside table causing corro-
sion and loss of function.
Summary of Contents for Amsco 3085 SP
Page 1: ...MAINTENANCE MANUAL Amsco 3085 SP Surgical Table After S N B420702 025 10 07 10 P764332 897 ...
Page 3: ...OPERATOR MANUAL Amsco 3085 SP Surgical Table 08 10 11 P150830 353 ...
Page 56: ......
Page 132: ......
Page 155: ...18 764332 899 Figure 5 AC Plate Assembly 1 7 3 4 5 6 2 9 10 P17 ...
Page 161: ...24 764332 899 Figure 8 Column Assembly 1 2 3 4 5 6 7 8 9 ...
Page 176: ...40 764332 899 Figure 15 Leg Section Assemblies 1 7 2 4 3 5 6 ...
Page 204: ......
Page 205: ......
Page 206: ......
Page 207: ......
Page 208: ......
Page 209: ......
Page 210: ......
Page 211: ......
Page 212: ......
Page 213: ......
Page 214: ......
Page 215: ......
Page 216: ......
Page 217: ......
Page 218: ......
Page 219: ......
Page 220: ......
Page 221: ......
Page 222: ......
Page 223: ......
Page 224: ......
Page 225: ......
Page 226: ......
Page 227: ......
Page 228: ......