4-5
Operating Instructions
Operator Manual
150830-353
2) When a normal patient load exceeds 700 lb (318 kg), moving the table
from an extreme Right Tilt may require the tabletops be level. When normal
patient load exceeds 900 lb (408 kg), moving the table from an extreme
Right Tilt may be slow or not operate.
3) When a reversed patient load exceeds 400 lb (181 kg), certain articula-
tions may be much slower than with a lighter load; for example, Reversed
Trendelenburg articulating. Additionally, when using X-ray tops with
bariatric patients, the automatic return-to-level function may not respond
until first initiating an independent articulation.
An optional foot control assembly is available for use in conjunction with the
hand control. See Figure 4-4 for identification of foot control functions.
NOTE: Battery-powered tables should be switched OFF after each procedure to
prevent unnecessary battery discharge. If a low battery condition is indicated by
the hand control BATTERY DOWN LED, refer to S
ECTION
6 for Battery Charging
Procedures.
1.
The following must be completed before any foot control
positioning functions are operable:
•
Hand control connected.
•
Control turned ON.
•
Floor locks engaged.
•
ORIENT PATIENT button activated (green LED ON) to indicate patient's
position on table (see Figures 4-2 and 4-3).
4.2.2 Optional Foot
Control Operation
WARNING – PERSONAL
INJURY HAZARD:
•
Unanticipated table move-
ment could cause patient in-
jury. Patient must be secured
to the table in accordance
with recommended position-
ing practices.
•
Do not immerse any part of
foot control in liquids; this
could cause unanticipated
table movement, leading to
patient injury. Always cover
control with a plastic bag
before using.
CAUTION – POSSIBLE
EQUIPMENT DAMAGE: Route
the hand control cord (and
optional HERMES
®
-Ready
1
or
ACT Enabled™ interface cord
and/or optional foot control
cord, if applicable) clear of
any pinch points where the
cord(s) could be damaged.
TRENDELENBURG
SIDE TILT
HEIGHT
Figure 4-4. Foot Control
Trendelenburg
Side Tilt to Left
Lower Height
Reverse
Trendelenburg
Raise
Height
Side Tilt to Right
TREND
REVERSE
RAISE
LOWER
LEFT
RIGHT
(International Symbols)
(USA/Canadian Controls)
1
HERMES-Ready is a registered trade-
mark of Computer Motion.
Summary of Contents for Amsco 3085 SP
Page 1: ...MAINTENANCE MANUAL Amsco 3085 SP Surgical Table After S N B420702 025 10 07 10 P764332 897 ...
Page 3: ...OPERATOR MANUAL Amsco 3085 SP Surgical Table 08 10 11 P150830 353 ...
Page 56: ......
Page 132: ......
Page 155: ...18 764332 899 Figure 5 AC Plate Assembly 1 7 3 4 5 6 2 9 10 P17 ...
Page 161: ...24 764332 899 Figure 8 Column Assembly 1 2 3 4 5 6 7 8 9 ...
Page 176: ...40 764332 899 Figure 15 Leg Section Assemblies 1 7 2 4 3 5 6 ...
Page 204: ......
Page 205: ......
Page 206: ......
Page 207: ......
Page 208: ......
Page 209: ......
Page 210: ......
Page 211: ......
Page 212: ......
Page 213: ......
Page 214: ......
Page 215: ......
Page 216: ......
Page 217: ......
Page 218: ......
Page 219: ......
Page 220: ......
Page 221: ......
Page 222: ......
Page 223: ......
Page 224: ......
Page 225: ......
Page 226: ......
Page 227: ......
Page 228: ......