4-8
150830-353
Operator Manual
Operating Instructions
4.3 Optional
Operating Room
Control System
(ORCS) Operation
WARNING – PERSONAL
INJURY HAZARD: Unantici-
pated table movement could
cause patient injury. Patient
must be secured to the table
in accordance with recom-
mended positioning prac-
tices.
CAUTION – POSSIBLE
EQUIPMENT DAMAGE:
Route the hand control cord
(and optional HERMES
®
-
Ready
1
or ACT Enabled™ in-
terface cord and/or optional
foot control cord, if appli-
cable) clear of any pinch
points where the cord(s)
could be damaged.
Figure 4-6. HERMES-Ready System Connection
4.3.1 HERMES-Ready
System Operation
1
HERMES-Ready is a registered trade-
mark of Computer Motion.
Align Red
Dots When
Connecting
HERMES-
Ready System
Interface
Cable
Black Receptacle (HERMES
Interface Cable Connection)
Black Strain Relief
For voice-activation of Amsco 3085 SP Surgical Table functions, either the
HERMES
®
-Ready
1
or ACT Enabled™ Interface System and appropriate hand
control are needed.
NOTE: Battery-powered tables should be switched OFF after each procedure
to prevent unnecessary battery discharge. If a low battery condition is indicated
by the hand control BATTERY DOWN LED, refer to S
ECTION
6 for Battery
Charging Procedures.
A HERMES-Ready Interface System and a HERMES-Ready Hand Control are
required for voice-activation of the HERMES-Ready Amsco 3085 SP table
functions.
IMPORTANT:
Use the HERMES-Ready 3085 SP hand control with the
blue
strain relief.The standard 3085 SP hand control with a
red
strain relief tail on the
connector will not connect to the HERMES-Ready table. For proper HERMES
System operation,
ensure HERMES-Ready table is interfaced with the HERMES
System only, not with any other Operating Room Control System.
NOTE:
If the HERMES Interface System and the table hand control are actuated
simultaneously, the hand control has priority.
1.
The following must be completed before any HERMES Voice-
Activated positioning functions are operable:
• Hand control connected (HERMES-Ready unit with blue tail on connection).
• Control turned ON.
• Floor locks engaged.
• ORIENT PATIENT button activated (green LED ON) to indicate patient's
position on table (see Figures 4-2 and 4-3).
2. Connect HERMES Interface System to HERMES-Ready 3085 SP table by
aligning HERMES interface cord black plug red dot with red dot of black
receptacle on HERMES-Ready table and pushing plug in to complete
connection (see Figure 4-6).
NOTE: A spring-loaded lock ring locks the plug into the receptacle. When
disconnecting the interface system, pull back on the lock ring before pulling
the plug from receptacle.
3. For operating instructions, refer to
HERMES Operating Room Control
Center Operating and Maintenance Manual and Appendix
provided with
HERMES System.
NOTE: When the HERMES System is powered up and it initially interrogates
the table, the HERMES display screen should acknowledge it as the table.
Summary of Contents for Amsco 3085 SP
Page 1: ...MAINTENANCE MANUAL Amsco 3085 SP Surgical Table After S N B420702 025 10 07 10 P764332 897 ...
Page 3: ...OPERATOR MANUAL Amsco 3085 SP Surgical Table 08 10 11 P150830 353 ...
Page 56: ......
Page 132: ......
Page 155: ...18 764332 899 Figure 5 AC Plate Assembly 1 7 3 4 5 6 2 9 10 P17 ...
Page 161: ...24 764332 899 Figure 8 Column Assembly 1 2 3 4 5 6 7 8 9 ...
Page 176: ...40 764332 899 Figure 15 Leg Section Assemblies 1 7 2 4 3 5 6 ...
Page 204: ......
Page 205: ......
Page 206: ......
Page 207: ......
Page 208: ......
Page 209: ......
Page 210: ......
Page 211: ......
Page 212: ......
Page 213: ......
Page 214: ......
Page 215: ......
Page 216: ......
Page 217: ......
Page 218: ......
Page 219: ......
Page 220: ......
Page 221: ......
Page 222: ......
Page 223: ......
Page 224: ......
Page 225: ......
Page 226: ......
Page 227: ......
Page 228: ......