4-9
Operating Instructions
Operator Manual
150830-353
WARNING – PERSONAL
INJURY HAZARD: Unantici-
pated table movement could
cause patient injury. Patient
must be secured to the table
in accordance with recom-
mended positioning prac-
tices.
CAUTION – POSSIBLE
EQUIPMENT DAMAGE:
Route the hand control cord
(and optional HERMES
®
-
Ready
1
or ACT Enabled™ in-
terface cord and/or optional
foot control cord, if appli-
cable) clear of any pinch
points where the cord(s)
could be damaged.
Figure 4-7. ACT Enabled System Connection
Align Red Dots
When Connect-
ing ACT Enabled
System Interface
Cable
Black Receptacle (ACT Enabled
Interface Cable Connection)
Black Strain Relief
4.3.2 ACT Enabled
System Operation
1
HERMES-Ready is a registered trademark of Computer Motion.
An ORCS (not provided by STERIS) and an ACT Enabled Hand Control are
required for voice- and/or touch panel-activation of the ACT Enabled Amsco
3085 SP table functions.
NOTE: Battery-powered tables should be switched OFF after each procedure
to prevent unnecessary battery discharge. If a low battery condition is indicated
by the hand control BATTERY DOWN LED, refer to S
ECTION
6 for Battery
Charging Procedures.
IMPORTANT:
Use the ACT Enabled 3085 SP hand control with the
blue
strain
relief.The standard 3085 SP hand control with a
red
strain relief tail on the
connector will not connect to the ACT Enabled table. For proper ACT voice/touch
panel operation,
ensure ACT Enabled table is interfaced with the appropriate
Operating Room Control System. ACT Enabled tables can NOT be controlled
by the HERMES system.
NOTE:
If the ORCS and the table hand control are actuated simultaneously, the
hand control has priority.
1.
The following must be completed before any ACT Enabled
Voice-Activated positioning functions are operable:
• Hand control connected (ACT Enabled unit with blue tail on connection).
• Control turned ON.
• Floor locks engaged.
• ORIENT PATIENT button activated (green LED ON) to indicate patient's
position on table (see Figures 4-2 and 4-3).
2. Connect ORCS to ACT Enabled 3085 SP table by aligning ACT Enabled
interface cord black plug red dot with red dot of black receptacle on ACT
Enabled table and pushing plug in to complete connection (see Figure 4-7).
NOTE: A spring-loaded lock ring locks the plug into the receptacle. When
disconnecting the interface system, pull back on the lock ring before pulling
the plug from receptacle.
3. For operating instructions, refer to
ACT Enabled Operating Room Control
System Operating Manual
provided with that system.
Summary of Contents for Amsco 3085 SP
Page 1: ...MAINTENANCE MANUAL Amsco 3085 SP Surgical Table After S N B420702 025 10 07 10 P764332 897 ...
Page 3: ...OPERATOR MANUAL Amsco 3085 SP Surgical Table 08 10 11 P150830 353 ...
Page 56: ......
Page 132: ......
Page 155: ...18 764332 899 Figure 5 AC Plate Assembly 1 7 3 4 5 6 2 9 10 P17 ...
Page 161: ...24 764332 899 Figure 8 Column Assembly 1 2 3 4 5 6 7 8 9 ...
Page 176: ...40 764332 899 Figure 15 Leg Section Assemblies 1 7 2 4 3 5 6 ...
Page 204: ......
Page 205: ......
Page 206: ......
Page 207: ......
Page 208: ......
Page 209: ......
Page 210: ......
Page 211: ......
Page 212: ......
Page 213: ......
Page 214: ......
Page 215: ......
Page 216: ......
Page 217: ......
Page 218: ......
Page 219: ......
Page 220: ......
Page 221: ......
Page 222: ......
Page 223: ......
Page 224: ......
Page 225: ......
Page 226: ......
Page 227: ......
Page 228: ......