6-6
150830-353
Operator Manual
Routine Maintenance
6.5 Battery Charging
Procedure
6.5.1 Electric-Powered
Tables
Control batteries* are recharged automatically when table is being used and do
not require additional charging.
Motor and control batteries* will require recharging on a periodic basis
depending on frequency of table usage. Low or discharged battery conditions
are indicated by LEDs on the hand control as explained in the Hand Control
Diagnostic Chart in
S
ECTION
7
.
Lead acid batteries last longer if NOT fully discharged. Therefore, to obtain the
longest life and capacity from your 3085 batteries, always connect ac power
cord to table base and plug into an appropriate ac receptacle as often as
possible, and as long as possible. If this is not always possible, recharge
batteries at the following times:
•
When table is first put into service.
•
Every two weeks when table is in normal service; more often if usage
demands.
•
Whenever a low battery indicator LED is ON.
•
If table remains in extended storage, batteries must be charged every six
months.
•
It is
not necessary
to have
hand control
ON to charge batteries.
•
On either electric-powered or battery-powered 3085 SP, power cord may be
left plugged into appropriate receptacle indefinitely. It will not harm table nor
table batteries.
NOTE: If batteries will not charge, refer to S
ECTION
7 for possible causes and
corrective actions.
Recharge batteries as follows:
1. Connect ac power cord to table base and plug into an appropriate ac
receptacle (see Figure 6-3).
6.5.2 Battery-Powered
Tables
WARNING – PERSONAL
INJURY AND/OR EQUIPMENT
DAMAGE HAZARD: Repairs
and adjustments to this equip-
ment must be made only by
fully qualified service person-
nel. Nonroutine maintenance
performed by inexperienced,
unqualified personnel or
installation of unauthorized
parts could cause personal
injury, invalidate the war-
ranty, or result in costly dam-
age. Contact STERIS regard-
ing service options.
WARNING – TRIPPING
HAZARD: Route the power
cord to the receptacle in a
position so it will not be
tripped over by personnel in
the area.
* All motor and control batteries are a sealed, lead-acid gel electrolyte-type, with a
nominal life of four years.
Connect AC Power Cord
Cover Raised
Figure 6-3. Change Batteries
Summary of Contents for Amsco 3085 SP
Page 1: ...MAINTENANCE MANUAL Amsco 3085 SP Surgical Table After S N B420702 025 10 07 10 P764332 897 ...
Page 3: ...OPERATOR MANUAL Amsco 3085 SP Surgical Table 08 10 11 P150830 353 ...
Page 56: ......
Page 132: ......
Page 155: ...18 764332 899 Figure 5 AC Plate Assembly 1 7 3 4 5 6 2 9 10 P17 ...
Page 161: ...24 764332 899 Figure 8 Column Assembly 1 2 3 4 5 6 7 8 9 ...
Page 176: ...40 764332 899 Figure 15 Leg Section Assemblies 1 7 2 4 3 5 6 ...
Page 204: ......
Page 205: ......
Page 206: ......
Page 207: ......
Page 208: ......
Page 209: ......
Page 210: ......
Page 211: ......
Page 212: ......
Page 213: ......
Page 214: ......
Page 215: ......
Page 216: ......
Page 217: ......
Page 218: ......
Page 219: ......
Page 220: ......
Page 221: ......
Page 222: ......
Page 223: ......
Page 224: ......
Page 225: ......
Page 226: ......
Page 227: ......
Page 228: ......