6-7
Routine Maintenance
Operator Manual
150830-353
2. Allow a
minimum of 48 hours
for full battery charge. See chart below:
Charging Time
Portion Charge
24 hours
90 percent
36 hours
95 percent
48 hours
100 percent
3. Verify low battery indicator LED is off and disconnect ac power.
4. As an alternative to
Step 2
, battery table may be used during daytime on
battery and charged at night for six to eight hours charging time minimum.
OR
Table may be used during week and charged over weekend (38 hours).
5. Depending on how batteries are cared for, table battery life can vary greatly.
The following can shorten the battery's useful life:
• continuing to leave this type of battery discharged for a long period, or
• continuing to use table battery even though battery light is blinking and not
plugging table into appropriate receptacle, or
• not fully charging battery.
Summary of Contents for Amsco 3085 SP
Page 1: ...MAINTENANCE MANUAL Amsco 3085 SP Surgical Table After S N B420702 025 10 07 10 P764332 897 ...
Page 3: ...OPERATOR MANUAL Amsco 3085 SP Surgical Table 08 10 11 P150830 353 ...
Page 56: ......
Page 132: ......
Page 155: ...18 764332 899 Figure 5 AC Plate Assembly 1 7 3 4 5 6 2 9 10 P17 ...
Page 161: ...24 764332 899 Figure 8 Column Assembly 1 2 3 4 5 6 7 8 9 ...
Page 176: ...40 764332 899 Figure 15 Leg Section Assemblies 1 7 2 4 3 5 6 ...
Page 204: ......
Page 205: ......
Page 206: ......
Page 207: ......
Page 208: ......
Page 209: ......
Page 210: ......
Page 211: ......
Page 212: ......
Page 213: ......
Page 214: ......
Page 215: ......
Page 216: ......
Page 217: ......
Page 218: ......
Page 219: ......
Page 220: ......
Page 221: ......
Page 222: ......
Page 223: ......
Page 224: ......
Page 225: ......
Page 226: ......
Page 227: ......
Page 228: ......