4-7
Operating Instructions
Operator Manual
150830-353
•
Lower Height
– 27" (686 mm) minimum.
Depress right side of HEIGHT pedal (located in the center position of foot
control pedals) and release pedal when desired position has been reached
to automatically stop tabletop and lock it in position.
•
Side Tilt to Left
– 18° maximum from horizontal.
Depress left side of SIDE TILT pedal (located in the right position of foot
control pedals) and release pedal when desired position has been
reached to automatically stop tabletop and lock it in position.
•
Side Tilt to Right
– 18° maximum from horizontal.
Depress right side of SIDE TILT pedal (located in the right position of foot
control pedals) and release pedal when desired position has been
reached to automatically stop tabletop and lock it in position.
NOTE: Momentary delay may occur when activating side tilt while the
safety mechanism disengages tilt-lock function.
IMPORTANT:
For table positioning, when patient load exceeds average
weights, note the following:
1) When a normal patient load exceeds 700 lb (318 kg), Reflex and Return-
to-Level articulations may be slow or not operate. Use other articulations
to move the tabletops to the desired position.
2) When a normal patient load exceeds 700 lb (318 kg), moving the table
from an extreme Right Tilt may require the tabletops be level. When normal
patient load exceeds 900 lb (408 kg), moving the table from an extreme
Right Tilt may be slow or not operate.
3) When reversed patient load exceeds 400 lb (181 kg), certain articulations
may be much slower than with lighter loads; for example, Reversed
Trendelenburg articulation.
When not in use both during and between procedures, the hand control should
be attached to the table side rail.
The foot control, if used, should be bagged and placed on the floor near the
surgical area. When the foot control is not required for a procedure, it should be
unplugged from the table and stored with other accessories until needed. Never
store the foot control (or any other objects) on the table base.
4.2.3 Care of Controls
When Not In Use
CAUTION – POSSIBLE
EQUIPMENT DAMAGE: Hang
the hand control from the side
rail (or end rail) of the table
when not in use, to avoid
possible damage to the con-
trol.
Summary of Contents for Amsco 3085 SP
Page 1: ...MAINTENANCE MANUAL Amsco 3085 SP Surgical Table After S N B420702 025 10 07 10 P764332 897 ...
Page 3: ...OPERATOR MANUAL Amsco 3085 SP Surgical Table 08 10 11 P150830 353 ...
Page 56: ......
Page 132: ......
Page 155: ...18 764332 899 Figure 5 AC Plate Assembly 1 7 3 4 5 6 2 9 10 P17 ...
Page 161: ...24 764332 899 Figure 8 Column Assembly 1 2 3 4 5 6 7 8 9 ...
Page 176: ...40 764332 899 Figure 15 Leg Section Assemblies 1 7 2 4 3 5 6 ...
Page 204: ......
Page 205: ......
Page 206: ......
Page 207: ......
Page 208: ......
Page 209: ......
Page 210: ......
Page 211: ......
Page 212: ......
Page 213: ......
Page 214: ......
Page 215: ......
Page 216: ......
Page 217: ......
Page 218: ......
Page 219: ......
Page 220: ......
Page 221: ......
Page 222: ......
Page 223: ......
Page 224: ......
Page 225: ......
Page 226: ......
Page 227: ......
Page 228: ......