4-1
Operating Instructions
Operator Manual
150830-353
4
4.1 Attach Headrest
and Orient Patient
OPERATING INSTRUCTIONS
For maximum patient positioning flexibility, the Amsco
®
3085 SP Surgical Table
is designed so the headrest can be attached to either end of the table.
IMPORTANT:
Control must be oriented as to the patient's position on table
before any positioning functions are operable. When the table is turned on with
the hand control, it will automatically activate in NORMAL patient orientation.
The user can then select REVERSE orientation if desired.
NOTE: Thumbscrews located under tabletop frame must be loosened before
headrest can be attached or removed.
1. Determine desired patient position and attach headrest to table end to obtain
this desired position (see Figure 4-1) as follows:
NOTE: The headrest (head section) is intended only to support the patient head
or feet. Load rating is 77 lb (35 kg).
a. Insert rods extending from each side of headrest attachment into bores
provided in either end of table frame.
b. Reach under tabletop frame and fully tighten both thumbscrews (one on
each side of frame) to secure headrest attachment in place. Refer to
S
ECTION
4.4, H
EADREST
P
OSITIONING
,
for adjustment procedures.
2. Verify power is ON and table floor locks are properly engaged.
NOTE: If the wrong function selection button is accidentally pressed, press the
correct function button to override the incorrect selection.
WARNING – EXPLOSION
HAZARD: Table must not be
used in the presence of flam-
mable anesthetics.
WARNING – TIPPING
HAZARD:
•
Do not place patient on the
table unless floor locks are
engaged.
•
Do not release floor locks
while patient is on table.
WARNING – PERSONAL
INJURY HAZARD: If the integ-
rity of the external protective
earth conductor installation
or arrangement is in doubt,
operate the table from its in-
ternal power source.
Attach Headrest
Tighten Thumbscrews
Figure 4-1. Attach Headrest
Summary of Contents for Amsco 3085 SP
Page 1: ...MAINTENANCE MANUAL Amsco 3085 SP Surgical Table After S N B420702 025 10 07 10 P764332 897 ...
Page 3: ...OPERATOR MANUAL Amsco 3085 SP Surgical Table 08 10 11 P150830 353 ...
Page 56: ......
Page 132: ......
Page 155: ...18 764332 899 Figure 5 AC Plate Assembly 1 7 3 4 5 6 2 9 10 P17 ...
Page 161: ...24 764332 899 Figure 8 Column Assembly 1 2 3 4 5 6 7 8 9 ...
Page 176: ...40 764332 899 Figure 15 Leg Section Assemblies 1 7 2 4 3 5 6 ...
Page 204: ......
Page 205: ......
Page 206: ......
Page 207: ......
Page 208: ......
Page 209: ......
Page 210: ......
Page 211: ......
Page 212: ......
Page 213: ......
Page 214: ......
Page 215: ......
Page 216: ......
Page 217: ......
Page 218: ......
Page 219: ......
Page 220: ......
Page 221: ......
Page 222: ......
Page 223: ......
Page 224: ......
Page 225: ......
Page 226: ......
Page 227: ......
Page 228: ......