7-10
DSR-80/80P/60/60P
7-5. TAPE PATH ADJUSTMENT
(TAPE PATH FINE ADJUSTMENTS AT ENTRANCE AND EXIT SIDES)
Required tools
Alignment tape XH2-1AST
: 8-967-999-02
Tape guide adjustment driver
: J-6440-850-A
Nutdriver (width across flat 4.5 mm) : 7-700-751-01
Dual trace oscilloscope
Check procedure
1. Connect an oscilloscope as follows :
CH-1 :
TP104/RP-103 board (D-1)
TP302/RP-101 board (D-1)
CH-2 : TP207/SV-184 board (F-1)
TRIG : CH-2
2. Set the alignment cassette XH2-1AST (standard
cassette) on the VTR and place a weight on a
cassette so that a cassette is securely fixed in
position.
3. Enter the PLAY mode.
4. Adjust the Variable VOLTS/DIV control of an
oscilloscope so that the maximum amplitude of
the RF waveform becomes the three DIVISIONS
sharp on an oscilloscope.
5. Turn the TG-3 height adjustment nut and TG-6
top flange counter-clockwise once so that the top
flanges are clear of tape, then turn them
clockwise until the RF waveform shown in
Fig.-1 is obtained.
Repeat the operation from STOP
→
PLAY
several times and confirm that the waveform
shown in Fig.-1 is obtained always. Then fix the
TG-6 top flange.
Note :
.....
Do not press TG-6 strongly
downward when fixing it.
.....
End the TG-3 and TG-6
adjustments with clockwise
rotation. (Do not end the
adjustment with counter-clockwise
rotation.)
DSR-80/80P
DSR-60/60P
5 Div
Peak of waveform
max
3 Div
Fig.-1