Operational procedures
35
850-164692 / Rev.B
Previously recorded data
The
Calibration
program also allows you to work with
previously recorded calibration data.
If you have used the ER60 to record raw data from a previous
calibration session you may operate the ER60 in replay mode
and perform the described calibration procedure as if the ER60
was operating in
Normal
mode. When operating in
Replay
mode you can set the ping rate to maximum since this will not
affect the data quality.
Offline calibration
The calibration program may be run offline without an
operational ER60. To start it, locate the file calibration.exe. If
you have performed a standard installation, this file is normally
found in the following path:
c:\program files\simrad\scientific\ek60\bin\calibration.exe
If you have a calibration file previously generated by the ER60
calibration program, you can import the file in the calibration
program.
1
Open the
File
menu in the calibration window, choose
Open.
2
Select the file you would like to import, and click
OK
.
3
Examine and edit the previously recorded calibration data.
Edited data may be saved and you can update the transducer
parameters using the imported or edited calibration data. This
also enables you to use different calibration files from different
environmental conditions.
Related topics
→
Environment, page 77
→
Normal Operation, page 102
→
Ping Control, page 105
→
Single Target Detection, page 117