Simrad ER60 Operator manual
30
850-164692 / Rev.B
The same structure can be repeated for the different pulse
durations for each frequency. When you save the calibration file,
create a file name that includes the current date and the current
output power. Save the information as a TXT file. Example:
....\0.256msec\cal010603-2000w.txt
While the calibration takes place, we recommend that you also
collect a raw data file. Browse the file directory in the Store
dialogue box to locate the correct directory. The raw data file
name will be generated automatically, but you may specify a
prefix. The raw data file is used for replay.
The same procedure can be repeated for each of the installed
frequencies.
Observe the following calibration procedure.
1
Ensure that only one transceiver is installed in the ER60
echo sounder software.
2
From the
Operation
menu, select
Ping control
.
3
In the
Ping Control
dialogue, set
Ping rate
to
Interval
and 1 second.
- This can also be done from the
Toolbar
.
4
From the
Operation
menu, select
Normal.
5
In the
Normal Operation
dialogue, select transceiver, and
switch to
Active
mode.
6
In the
Normal Operation
dialogue, set the
Transmit
Power
to the level you wish to calibrate.
7
In the
Normal Operation
dialogue, choose the
Pulse
Duration
you wish to calibrate.
8
Right-click in an echogram, select
Range
on the shortcut
menu, and set the range for one of the echogram views to
cover the range you wish to see.
- This range should include the depth range where you
expect to find the reference target.
9
Check that you see the reference target in the
Single Echo
view.
10
Click the right mouse button in the
Single Echo
view
corresponding to the echogram to open the
Single Target
Detection
dialogue box.
11
In the
Single Target Detection
dialogue, click the
Calibration
button to start the calibration program, and to
create a new
Calibration
window.
- The calibration program allows you to record new
calibration data, or read previously recorded calibration
data.