PowerSync Analyzer Reference Manual
March 9, 2010
Sifos Technologies page
74
4.12.2.
Measure Peak Detection Voltage on Port 1,2
This script will capture the peak detection voltage measured while the detection signature is invalid.
If these commands are sequentially executed in a Wish Shell, it will appear as follows:
# Set up a 34K ohm (out-of-range) detection signature and configure port for
# ALT A, MDI-X PSE. Disconnect any power.
alt
1,2 a
polarity
neg
psa_disconnect
passive
r 34 c 0
# Setup the triggering to a detection pulse – note that this assumes
# detection pulses return to approximately zero volts.
trig1
rising level 2 arm
# Configure and initiate the peak voltage measurement.
vdcpeak
trig on max period 500m timeout 10
port
connect
set status [
psa_wait
vdcpeak
]
set peak_result [lindex $status 3]