PowerSync Analyzer Reference Manual
March 9, 2010
Sifos Technologies page
63
Command
Port
Command Parameters
Query
Returned
Parameters
<
tr
ig
off
|
on
|
ext
> <
p
eriod
interval
> <
ti
meout
10
| 100>
READY | ARMED |
MEASURING |
TIMEOUT
stat
List of 100 Voltage
Values
TRIG
OFF
| ON | EXT
MAX
| MIN
Sampling Period
vdct
race
<
port
>
Configures and collects a DC Voltage Trace of 256
samples across the PoE bus.
tr
ig
off
enables immediate execute of DC
measurement when “stat” query is issued.
tr
ig
on
specifies that the measurement will be
performed when the trigger condition specified on
Trig1 is achieved. Trig1 will be armed
automatically when “stat” query is issued.
tr
ig
ext
specifies that the measurement will be
performed when the external (event) trigger is
asserted.
p
eriod
specifies that an sampling period will be
defined.
interval
specifies the sampling period as “Nm” where
N is 10, 20, 50, 100, 200, 500 mSec or “Ns”
where N is 1, 2, 5, or 10 seconds. Default value
is 100 mSec.
ti
meout
10
specifies that a hardware triggered
measurement should time out if no trigger is
found within 10 seconds following trigger arm.
ti
meou
t
100
specifies that a hardware triggered
measurement should time out if no trigger is
found within 100 seconds following trigger arm.
?
Timeout
10
| 100
<
tr
ig
off
|
on
|
ext
> <
p
eriod
interval
> <
ti
meout
10
| 100>
READY | ARMED |
MEASURING |
TIMEOUT
stat
Average DC Current
TRIG
OFF
| ON | EXT
Averaging Period
idca
verage
<
port
>
Configures and performs an Average DC Current
measurement on a PoE Pair.
tr
ig
off
enables immediate execute of DC
measurement when “stat” query is issued.
tr
ig
on
specifies that the measurement will be
performed when the trigger condition specified on
Trig1 is achieved. Trig1 will be armed
automatically when “stat” query is issued.
tr
ig
ext
specifies that the measurement will be
performed when the external (event) trigger is
asserted.
p
eriod
specifies that an averaging period will be
defined.
interval
specifies the averaging period as “Nm” where
N is 10, 20, 50, 100, 200, 500 mSec or “Ns”
where N is 1, 2, 5, or 10 seconds. Default value
is 100 mSec.
ti
meout
10
specifies that a hardware triggered
measurement should time out if no trigger is
found within 10 seconds following trigger arm.
ti
meou
t
100
specifies that a hardware triggered
measurement should time out if no trigger is
found within 100 seconds following trigger arm.
?
Timeout
10
| 100
Summary of Contents for PowerSync PSA100
Page 26: ...PowerSync Analyzer Reference Manual March 9 2010 Sifos Technologies page 26 ...
Page 56: ...PowerSync Analyzer Reference Manual March 9 2010 Sifos Technologies page 56 ...
Page 84: ...PowerSync Analyzer Reference Manual March 9 2010 Sifos Technologies page 84 ...
Page 110: ...PowerSync Analyzer Reference Manual March 9 2010 Sifos Technologies page 110 ...
Page 120: ...PowerSync Analyzer Reference Manual March 9 2010 Sifos Technologies page 120 ...