PowerSync Analyzer Reference Manual
March 9, 2010
Sifos Technologies page
60
4.6.
Test Port Configuration Commands
The following commands are directed to specific test ports and generally used for static configuration of test resources
at that port. Whenever the
port
parameter is provided, the default slot-port for that and all subsequent commands will
be set as will the command prompt.
Command
Port
Command Parameters
Query
Returned
Parameters
a
|
b
alt
<
port
>
Configures Port Mux for powered pairs. Port will default
to Alt A when PSA chassis is first powered on.
a
Select ALT A pairs
b
Select ALT B pairs
?
A
| B
<
p
os
|
n
eg
>
pol
arity
<
port
>
Configures Bus Switch for power polarity. Blade will
default to negative polarity (MDI-X) when PSA chassis is
first powered on.
p
os
sets to PoE Positive Polarity (MDI)
n
eg
sets to PoE Negative Polarity (MDI-X)
?
POS |
NEG
<
c
onnect
<
tr
igout
>
|
i
solate
<
tr
igout
>>
port
<
port
>
Configures Port Isolation Switch. Isolation removes
detection passives and AC MPS load resistance from
port. Port will default to isolated when PSA chassis is first
powered on. Option provided to generate a trigger out on
each issuance of this command.
c
onnect
connects the port
i
solate
disconnects the port
tr
igout
: Generate hardware trigger on command
execution
?
Connected |
Isolated
<
r
resistance
>
<
c
capacitance
>
Resistance Value
passive
<
port
>
Configures passive detection entity. Port will default to
resistance = 24 K
Ω
, capacitance = .1
μ
F (“0”) when PSA
chassis is first powered on.
r
specifies a resistance setting
resistance
is resistance value in K
Ω
from 8 to 39
c
specifies a capacitance setting
capacitance
is capacitance value in
μ
F = 0, 5, 7, 11, 47,
52, 54, or 58. (Note Type 1 PSA blades support 0, 5,
7, and 11
μ
F.)
?
Capacitance Value
<
i
current
<
tr
igout
>>
stat
ON |
OFF
iload
<
port
>
Configures active load entity to a steady state current
load. Option provided to generate a trigger out on each
issuance of this command.
i
current
is current load in units of mA. Current load has
range of 0 to 511 mA with 0.5 mA steps.
(2 mA steps
on firmware versions prior to 1.4)
tr
igout
: Generate hardware trigger coincident with static
load change.specifies current entry
?
Current Value
<
0
|
1
|
2
|
3
|
4
>
Class Setting
class
<
port
>
Configures current load to emulate classification current
for the specified class.
0
= Class 0 (2.5 mA)
1
= Class 1 (10.5 mA)
2
= Class 2 (18.5 mA)
3
= Class 3 (28 mA)
4
= Class 4 (40 mA)
?
Summary of Contents for PowerSync PSA100
Page 26: ...PowerSync Analyzer Reference Manual March 9 2010 Sifos Technologies page 26 ...
Page 56: ...PowerSync Analyzer Reference Manual March 9 2010 Sifos Technologies page 56 ...
Page 84: ...PowerSync Analyzer Reference Manual March 9 2010 Sifos Technologies page 84 ...
Page 110: ...PowerSync Analyzer Reference Manual March 9 2010 Sifos Technologies page 110 ...
Page 120: ...PowerSync Analyzer Reference Manual March 9 2010 Sifos Technologies page 120 ...