
PowerSync Analyzer Reference Manual
March 9, 2010
Sifos Technologies page
31
Trigger1 may be “ARMED” by pressing the
ARM Trigger
button. This enables a trigger to be generated as soon as
the next qualifying signal event on the PoE bus occurs. At that time, the trigger will fire just once and will need to be
re-armed in order to fire again. All triggers in the PowerSync Analyzer test port are a one-shot events.
The
Clear Trigger
button will disarm a previously ARMed trigger where the trigger event has yet to occur, thus
preventing an internal trigger event.
The
Send Trigger
button generates an immediate External (or Event) Trigger that is also useful for initiating meter
measurements. As described in Section 2.3.2, External Triggers are exported to all other test ports in the chassis and
also to the TRIG connector on the chassis controller.
The
Copy Settings
button will open the Port Replication Menu (
see Section 3.1.4
) so that current port trigger settings
may be copied to one or more other ports.
3.1.6. Load
Configuration
A key function of each PowerSync Analyzer test port is to provide a controlled DC current load to the PSE port. The
Load Configuration
resource menu
(Figure 3.8)
offers capability to configure and control both a static DC load
current as well as a transient load current event. It also controls the ability to emulate selected classification power-up
sequences that involve the PSA1200 load subsystem.
As with other configuration and meter
resource meters, the
Load Configuration
menu offers selection of Slot-Port, chassis,
and the ability to replicate settings.
Static Load Current refers to a steady state
DC load current that will be present once
the PoE bus voltage exceeds 15.5 VDC.
This current will not exist below that
voltage. The load current may be typed
into the
Powered Load
entry field or
entered by the slider control. The
Apply
Load
button will then send this
configuration to the selected slot-port and
update the Load Status field. The
Apply
Load
button will annunciate via
red
text
that a change has been selected and must
be applied.
Static load current can range from 0 to 511
mA with a granularity of 0.5 mA. Entered
values will be automatically rounded to
the nearest multiple of 0.5 mA.
The
Mark Load
control is disabled with the PSA-1200 and is reserved for use with future PSA hardware.
Load transients always consist of two sequential load current “steps”, each of which can be programmed to a unique
load current value and step duration. Upon completion of a load current transient, load current will return to the
original static load current value unless the
Hold Step 2 Current
button is selected. In this case, the static DC load
current will be adjusted to the value of the second load transient step current.
The load current value for each transient step ranges from 0 to 511 mA with granularity of 0.5 mA. The duration of
each step may be specified from 200
μ
sec to 1000 mSec. The duration of the second load transient step will be ignored
if the
Hold Step 2 Current
setting is selected since step 2 load current will then become the new steady state load.
Load transients can either be immediately initiated or they can be initiated upon a trigger event. The
Trigger Mode
setting specifies what the initiating mechanism will be.
Immediate
selection will cause the load transient to be
configured and initiated as soon as the
Initiate Transient
button is pressed.
Waveform
selection means that the
Initiate Transient
button will configure the load current transient, but that the load transient will not occur until a
Figure 3.8
PSA Active Load Configuration Menu
Summary of Contents for PowerSync PSA100
Page 26: ...PowerSync Analyzer Reference Manual March 9 2010 Sifos Technologies page 26 ...
Page 56: ...PowerSync Analyzer Reference Manual March 9 2010 Sifos Technologies page 56 ...
Page 84: ...PowerSync Analyzer Reference Manual March 9 2010 Sifos Technologies page 84 ...
Page 110: ...PowerSync Analyzer Reference Manual March 9 2010 Sifos Technologies page 110 ...
Page 120: ...PowerSync Analyzer Reference Manual March 9 2010 Sifos Technologies page 120 ...