PowerSync Analyzer Reference Manual
March 9, 2010
Sifos Technologies page
71
4.10. PowerShell PSE Conformance Test Support Commands
PowerShell offers high level PSE characterization and analysis utilities for those systems that are enabled for PSE
Conformance Test Suite. These are described below.
Command
Slot
Command Parameters
Returned
Parameters
Backoff_Mode
get_sync_parms
<
port
>
Utility uses
psa_det_char
to analyze various aspects of PSE
detection signaling on the current port, then retains those
parameters into a global structure for use by PSE Conformance
Tests that require knowledge of detection signaling
characteristics. The utility also calculates a nominal sync delay
time for use with synchronization to open circuit detection
signaling. The array structure populated by
get_sync_parms
is:
psaPseChars(idleBackoff,$port)
psaPseChars(idleLevel,$port)
psaPseChars(idleMode,$port)
psaPseChars(idleRdet,$port)
psaPseChars(idleTrig,$port)
psaPseChars(ocDutyCycle,$port)
psaPseChars(syncEdge,$port)
psaPseChars(syncMode,$port)
psaPseChars(syncTrig,$port)
psaPseChars(syncWait,$port)
This array structure is volatile and will re-initialize anytime the
PSE Conformance Test Sequencer is run thus forcing new re-
characterization for every port tested.
1 = Analysis
Completed
psa_det_char
<
port
>
This utility performs a detailed analysis of PSE detection signaling
behaviors and is used by PSE Conformance Tests to adapt to the
variety of PSE technologies available. Characterizations is done
for both open circuit signaling and invalid signature response
signaling. Parameters returned are:
syncMode
: Open Circuit Signaling Type
syncTrig
: Optimal level for open circuit synchronization trigger
syncBackoff
: Minimim timing between open circuit pulses
ocDutyCycle
: Duty cycle of open circuit signaling above syncTrig
idleMode
: Idle State back-off type
idleTr
ig
: Optimal level for Idle State synchronization trigger
idleLevel
: Idle State floor voltage
idleBackoff
: Minimum timing between Idle state pulses
idleRdet
: Optimal detection signature for Idle State analysis
See Parameters…
4.11. PowerShell System & Chassis Commands
System commands are utilized to control and configure at the PowerSync Analyzer chassis level. The
psa
command
described in Section 4.5 above will define the chassis on which each of these commands will operate.
Command
Command Parameters
Returned
Parameters
<
-d
|
-s
>
psa_config
Inventories and reports status of available system resources (blades,
ports, firmware version). Inventory is performed upon currently selected
PSA chassis (see “psa” command). Default output is TCL List of
inventory.
-
d
Causes inventory to be displayed in tabular form.
-s
Returns a list of populated slots
{Port + Type +
Version}
Summary of Contents for PowerSync PSA100
Page 26: ...PowerSync Analyzer Reference Manual March 9 2010 Sifos Technologies page 26 ...
Page 56: ...PowerSync Analyzer Reference Manual March 9 2010 Sifos Technologies page 56 ...
Page 84: ...PowerSync Analyzer Reference Manual March 9 2010 Sifos Technologies page 84 ...
Page 110: ...PowerSync Analyzer Reference Manual March 9 2010 Sifos Technologies page 110 ...
Page 120: ...PowerSync Analyzer Reference Manual March 9 2010 Sifos Technologies page 120 ...