Power supply
12-24 V AC/DC power supply
Input voltage range
12/24 V AC/DC (10 ... 32 V DC)
Power consumption
2 VA
Isolation
Class II
Fuse
1000 mA T - Not replaceable
Short circuit protection
Module is protected from short circuit on the output con‐
nector. Both during mains and backup supply
Conformity
IEC 61010-1
OIML R 49-1
EMC:
EN 61000-6-3
EN 61000-6-2
115 to 230 V AC mains supply Input voltage range
115 to 230 V AC, +15% to −20%, 50-60 Hz
Power consumption
2 VA
Isolation
Class II
Fuse
250 mA T - Not replaceable
Short circuit protection
Module is protected from short circuit on the output con‐
nector. Both during mains and backup supply
Conformity
IEC 61010-1
OIML R49-1
EMC:
EN 61000-6-3
EN 61000-6-2
Input cable for
12/24 V AC/DC and
115 to 230 V AC power supply
Factory-mounted PUR cable
2 x 1 mm
2
(brown, blue)
Length = 3 m
Resistance
Sunlight and water
Outer diameter
7 mm (0.28")
Rated voltage
300 ... 500 V AC
Testing voltage
2000 V AC
Temperature range
Fixed laying:
Flexible application:
−40 to +90 °C (-40 to 194 °F)
−30 to +80 °C (-22 to 176 °F)
Bending radius
Min. 28 mm (fixed installation)
Pulling force
Max. 200 N
Output
Female connector
Backup battery
Male connector
1)
Lithium batteries are subject to special transportation regulations according to United Nations
"Regulation of Dangerous Goods, UN 3090 and UN 3091". Special transport documentation
is required to observe these regulations. This may influence both transport time and costs.
10.5
Modbus RTU
Siemens Flow Instruments Modbus RTU specification for add-on modules
Device type
Slave
Baud rates
1200, 2400, 4800, 9600, 19 200, 38 400 bits/sec.
Technical data
10.5 Modbus RTU
SITRANS MAG 8000
Operating Instructions, 11/2018, A5E03071515-AE
95