Flow rate unit (parameter 9)
Correction factor parameter 301
BBL42/h
22643.31867
BBL42/d
543439.6482
C.2
Parameter lists
Parameter lists
MAG 8000 is delivered with factory settings that are not stored as default values. Because
defaults values are not present in the meter, an automatic return to factory values is not
possible.
The default settings are available at www.siemens.com/flow (
).
Navigate to Tools & Downloads under MAG 8000.
Visible display information is indicated in the table by menu and index number. Remember to
enable displayed menus in parameter 130.
The abbreviations used in the display menu table are: Operator menu = O, Meter menu = M,
Service menu = Se, Data Logger menu = L, Statistic menu = St, Revenue menu = R.
Factory settings marked in
italic cannot be edited.
Note
Change of units
If the unit and factor is changed, all unit-related values become invalid.
It is recommended to reset all unit-related values.
C.2.1
1-99
Param‐
eter ID
Meter
version
Display
view
Parameter/data type
Factory settings
Data range
Fixed parameter or meter data that are not changeable
1
All
M1
Application identifier
Identity
Max. 15 characters.
Only numbers are visible in the
display (letters are shown as "_")
2
All
-
Application location
Location
Max. 16 characters
3
All
M3
Module type
MLFB dependent
MAG 8000 Basic
MAG 8000 Advanced
MAG 8000 CT Basic
MAG 8000 CT Advanced
MAG 8000 Irrigation
4
All
M4
Software version
x.xx
5
All
-
Sensor size
Sensor related
Max. 12 characters.
DN 25 to 1200 (1" to 48")
Appendix
C.2 Parameter lists
SITRANS MAG 8000
136
Operating Instructions, 11/2018, A5E03071515-AE