C.2.5
400-499
Param‐
eter ID
Meter
version
Display
view
Parameter/data type
Factory settings
Data range
Fixed parameter or meter data that are not changeable
400*
All
-
Output A enable
MLFB-dependent
Yes / No
401*
All
Se3
Pulse A direction
Forward
Forward, Reverse, Forward net,
Reverse net
402*
All
Se3
Amount per pulse A
Sensor-related
0 to 1x10
9
403*
All
-
Pulse width for pulse A
50 ms
10 ms, 25ms, 50 ms, 100 ms,
500 ms, 5 ms, 0.5 ms
404*
All
-
Output B enable
MLFB-dependent
Yes / No
405*
All
Se4
Pulse B function
Alarm
Pulse, Alarm, Call-up
406*
All
Pulse B direction
Reverse
Forward, Reverse, Forward net,
Reverse net
407*
All
Se4
Amount per pulse B
Sensor-related
0 to 1x10
9
408*
All
Pulse width for pulse B
50 ms
10 ms, 50 ms, 100 ms, 500 ms
420
All
M5
Device Communication Ad‐
dress
1
1 to 247
421**
All
M6
Baud rate
19 200
1200, 2400, 4800, 9600, 19 200,
38 400
422**
All
M7
Parity
Even 1 stop
Even 1 stop, Odd 1 stop, None
1 stop, None 2 stop,
423**
All
-
Interframe space
35
35 to 255
424**
All
-
Response delay
5
1 to 50 ms
425
All
-
Reset communication driver
No
Yes / No
This parameter is removed from
PDM parameter list. The func‐
tionality is implemented via
PDM menu "Device" → "Commu‐
nication Settings" → "Transfer"
*: The parameter is accessible via PDM menu "Device" →"Pulse guide offline".
**: The parameter is accessible via the PDM menu "Device" → "Communication settings".
C.2.6
500-599
Param‐
eter ID
Meter
version
Display
view
Parameter/data type
Factory settings
Data range
Fixed parameter or meter data that are not changeable
500
All
-
Latest service date
PS3 production date and
time
year-month-day T hours:mi‐
nutes:seconds
501
All
-
Operating hours since power-
up
0
hours
502
All
-
Battery operating time
0
hours
Appendix
C.2 Parameter lists
SITRANS MAG 8000
Operating Instructions, 11/2018, A5E03071515-AE