C.6
Features
Feature
MAG 8000 basic
MAG 8000 advanced
Measurement frequency
(battery power)
Max. 1/15 Hz
Max. 6.25 Hz
For MAG 8000 Irrigation
(7ME6880):
Max. 3.125 Hz
Totalizer
3
3
Pulse output
2, max. 50 Hz
2, max. 100 Hz
Communication
Add-on
Add-on
IrDA
Yes
Yes
Time and date
Yes
Yes
Data protection
Yes
Yes
Data logger
Yes
Yes
Application identifier
Yes
Yes
Alarm handling
Yes
Yes
Meter status
Yes
Yes
Diagnostics
Yes
Yes
Battery power management
Yes
Yes
Insulation test
-
Yes
Leakage detection
-
Yes
Meter utilization
-
Yes
Statistics
-
Yes
Tariff
-
Yes
Settle date (Revenue)
-
Yes
Features
Application identification (parameters 1 and 2)
Tag number (visible on display if numbers are selected) and meter location, up to 15 characters
per information.
Time and date (parameter 100)
Real time clock and date (max. 15 minutes drift per year)
Totalizer (parameters 101, 102 and 103)
● 3 totalizers: Forward, reverse, bidirectional netflow calculation and freely selectable start
value.
Note
CT version
The totalizers are not resettable for the CT version.
Appendix