92
Final procedures
MAMMOMAT Novation DR
SPB7-250.815.06.02.02
Siemens AG
12.05
CS SD 24
Page 92 of 116
Medical Solutions
Service PC and measuring instruments
0
Saving the configuration file
0
NOTE
Make sure that the backup floppy is inserted.
Main menu:
1. Select:
Configuration -> Save config file
.
2. Press
F2
to save the data.
Reset the exposure counter
Main menu:
1. Select:
Service -> Reset exposure counter
.
2. Delete the error memory with
Y
.
Deleting the error buffer
Main menu:
1. Select:
Service -> Delete error buffer
.
2. Delete the error memory with
Y
.
Saving the MAMMOMAT Parameters
Main menu
1. Select
Backup -> Copy Installation Area to Floppy -> Stand
.
2. Select
Backup -> Copy Installation Area to Floppy -> Panel
.
3. Exit the program with
F10
when the backup is completed.
NOTE
Keep the backup floppy with other documentation for the MAM-
MOMAT Novation
DR
in question.