38
Mechanical covers
MAMMOMAT Novation DR
SPB7-250.815.06.02.02
Siemens AG
12.05
CS SD 24
Page 38 of 116
Medical Solutions
5. Check the rotary motion of the swivel arm system by pressing the buttons
one by one.
The rotation of the swivel arm system should stop when the button is released or when
the end position is reached (
+135 degrees clockwise
and
-180 degrees
counterclockwise
). The projection angle is shown on the display at the lower part of
the stand.
Fig. 20: Display at the lower part of the stand
6. Set the projection angle to
60 degrees
with the presettings knob
.
The preset angle is displayed for approximately
3 s
.
7. Press one of the buttons
.
The rotary motion of the swivel arm system should stop automatically when the preset
projection angle (both + and -) is reached and at
0 degrees, +90 degrees or -90
degrees
.
NOTE
All eight rotary motion buttons (four on either side) must be
checked.
).
After the button has been released, the swivel arm system should
remain in the position set.
OPDIMA wing
1. Select the OPDIMA wing; OP is then displayed on the control console.
2. Set the projection angle to minimum with presetting knob
.
3. Check the rotary motion of the swivel arm system by pressing the buttons
one by one.
The rotation of the swivel arm system should stop when the button is released or when
the end position is reached (
+120 degrees clockwise
and
-120 degrees
counterclockwise
). The projection angle is shown on the display at the lower part of
the stand.