20
Power up the MAMMOMAT
MAMMOMAT Novation DR
SPB7-250.815.06.02.02
Siemens AG
12.05
CS SD 24
Page 20 of 116
Medical Solutions
Connections
0
Fig. 5:
Three incoming mains cables
1. Connect the incoming mains cable to the mains terminal (
) at the bottom
of the back of the stand.
2. Strain-relieve the incoming mains cable.
Power system Type of connection
Power voltage
Power connections
Single phase
Phase to neutral
208 V-277 V
Connect to
3
and
4
(see No.1
/
)
Two phase
Phase to phase
208 V-400 V phase to
phase.
Note: Max. 277 V
phase to neu-
tral/ground allowed.
Connect to
3
and
4
(see No.1
/
)
Use two phases (of three) in a nor-
mal power distribution sys-
tem.The leftover wires, neutral
and the third phase may be con-
nected to terminal
1
and
2
. See
.