
90
Final procedures
MAMMOMAT Novation DR
SPB7-250.815.06.02.02
Siemens AG
12.05
CS SD 24
Page 90 of 116
Medical Solutions
Settings date and time
0
MAMMOMAT Stand
0
Prerequisite
•
The correct date and time must be set in the service PC. If not, set the date at C:\> by
typing
date
<ENTER> and the time at C:\ by typing
time
<ENTER>
Test performance
1. Start the Service Software for the MAMMOMAT Stand on the Service PC.
2. In the
Main menu
, select
Configuration -> Clock
.
Fig. 61: MAMMOMAT Stand clock adjustment
3. Press
<F2>
.
With
<F2>
the date and time set in the service PC are taken over by the MAMMOMAT
stand.
BRICK
0
1. Call up the WH AWS service software and select
Utilities -> Escape to OS
. Enter then
the following command in the
Command
field.
start telnet brick
2. A Username and Password have to be entered
Username:
root
Password:
******