Checking dose settings
59
Page 59 of 116
Medical Solutions
3. Select the Procedure
QC-Raw
from the examination tab card.
Fig. 39:
select procedure
NOTE
Remember, you can always add or remove images to the "Proce-
dure" by right-clicking the mouse.
4. Select the appropriate exposure parameters according to the table
Tab. 12
Checking H & D dose settings for Mo/Mo
5. Position the dose meter probe 6 cm from the chest wall and 3 cm off the center of the
detector.
NOTE
If parts of the meter probe cover the AEC sensor, one will get a
higher dose.
Thus, make sure that the probe is not covering the AEC sensors.
This can be checked with the mounted compression plate. The
AEC sensors are drawn on the compression plate.
X
6. Start an exposure.
AEC
kV
Anode / Filter
Dose Value
Pixel Value
H
28 kV
Mo/Mo
D
28 kV
Mo/Mo