Collimator Adjustment
MAMMOMAT Novation DR
SPB7-250.815.06.02.02
Siemens AG
12.05
CS SD 24
Page 54 of 116
Medical Solutions
7. Use a magnification of
4
in the application software and go to the edges of the detector.
White areas should be visible at all edges of the image. This shows you that the X-ray
does not cover the entire detector and shows where the beam really ends.
8. Measure the white line thickness at each image edge with the application graphical
user interface and adjust the corresponding
XL
,
XR
,
YC
and
YS
values with the service
software.
X
9. Start an exposure.
10. Check if the white lines at the edges of the image are gone using a magnification of
4
in
the application graphical user interface.
11.
Save
the calibration and
Exit
from the Service Software
NOTE
If you see a white thin line on the chest wall when using the mag-
nification table, it might be the magnification table itself. Move the
magnification table out by approximately 1 mm and try again.
12. Record the measured values in the "Start-up Protocol" on
.