External exposure lamp (optional)
0
This should only be performed if the customer requires it.
The connection on the master board works as a shutter. The signal from the master board
is only active during exposure, thus the exposure lamp will only be lit when the signal is
active.
Connect the cable to the lamp holder in series with the power supply.
CAUTION
When connecting the exposure lamp to the master board, make
sure that the electrical load does not exceed 24VAC / 25W
¹
1. Connect the exposure lamp to the connector X759 located on the master board D750,
see
.
Fig. 4:
Connecting the exposure lamp
NOTE
Do not use power from the MAMMOMAT stand.