Image Quality
MAMMOMAT Novation DR
SPB7-250.815.06.02.02
Siemens AG
12.05
CS SD 24
Page 88 of 116
Medical Solutions
6. Select the local service menu of the
AWS -> Acquisition System -> Image Process-
ing
7. Make a copy of the PPG on which you based the optimization of the contrast and the
brightness, e.g. "Normal" and change the name to e.g. "Standard".
8. Change the parameter "Window Center Shift" and "Factor make window width wider" to
the new values displayed in the configuration tool and save the PPG
The new calculated values for
Factor make window width wider
and
Window Center
Shift
have to be entered in the corresponding fields.
Tab. 16
Settings for normal parameters
Default
Result
Factor make window
width wider
1.6
Window Center Shift
150