Collimator Adjustment
MAMMOMAT Novation DR
SPB7-250.815.06.02.02
Siemens AG
12.05
CS SD 24
Page 50 of 116
Medical Solutions
•
Moving the collimator
IN
is done by
DECREASING
the
YC
value and
INCREASING
the
YS
,
XL
, and
XR
values.
Collimator beam check and calibration
0
You have to perform the collimator beam check with the
detector wing
for all
focus &
anode
combinations.
1. Make sure that a patient is registered.
2. Select the
detector wing. dr
is then displayed on the control console.
3. Mount the Collimator Mounted Phantom.
4. Connect the compression simulator.
5. Select
25 kV
and
50 mAs
on the control console.
X
6. Start an exposure.
7. View the image with a magnification of 4.
8. Slide the magnification glass along the image edges of the image and check if white
lines are visible.
NOTE
If you do not see white lines at the edges of the image, the collima-
tor beam calibration is correct and you can continue with the "Col-
limator field light calibration" on
.
Otherwise continue with the "Collimator beam calibration" on
(Collimator beam calibration / p. 50)
Collimator beam calibration
Collimator for the collimator beam is carried out as follows:
1. Make sure that the
detector wing
is selected.
dr
is then displayed on the control con-
sole.
2. Connect the compression plate simulator and the collimator mounted plexi phantom.
3. Connect the service PC to the MAMMOMAT stand and start the service software.