
Siemens AG
SPB7-250.815.06.02.02
MAMMOMAT Novation DR
12.05
CS SD 24
Power up the MAMMOMAT
19
Page 19 of 116
Medical Solutions
Mains voltage
0
WARNING
Before the power for the MAMMOMAT is turned on, make sure that
the cable plugs (left and right of the system) for the magnification
table are not shorted.
¹
The mains input can be either 1-phase or 2-phase, 50 or 60 Hz.
Electrical power to mains input:
0
Mains voltage
1-phase:
or
2-phase:
208, 230, 240 and 277 V ±10%
208, 230, 240, 277 and 400 V ±10%
Mains Frequency: 50 or 60 Hz
Maximum load:
10kVA (momentary) and 0.8 kVA (long time load).
Fuse:
25 A (external) 208 -> 400 V.