Siemens AG
SPB7-250.815.06.02.02
MAMMOMAT Novation DR
12.05
CS SD 24
Collimator Adjustment
53
Page 53 of 116
Medical Solutions
P
13. Record the measured values in the
"Start-up Protocol"
on
.
Collimator Wing Difference Calibration
The collimator wing difference calibration has to be adjusted
NOTE
If no precise wing difference calculation is performed, or if you
have to start from scratch, you can start with the "large focus/mol-
ly" values that have been adjusted with the "Collimator beam cali-
bration" on
(Collimator beam calibration / p. 50)
.
1. Select the
OPDIMA wing
by moving the detector into the park position.
OP
is then dis-
played on the control console.
2. Connect the compression plate simulator and the collimator-mounted plexi phantom.
3. Start the service software on the service PC and select:
Configuration -> Collimator -> Calibration of wing difference
Fig. 35:
Collimator wing difference calibration
4. Enter the numbers in the table below:
Tab. 11
Wing difference
5. Narrow the X-ray field so that the edges of the detector will not be exposed to the X-ray
and white lines are visible in the image at the edges of the detector. To do this, de-
crease the YC value and increase the
YS
,
XR
and
XL
values by
5
numbers.
X
6. Make sure a patient is registered and start an exposure.
Focus and Anode
XL
XR
YC
YS
Wing difference